USER GUIDE
SCC-AI Series Isolated Analog
Input Modules
The SCC-AI Series isolated analog input modules (SCC-AIXX) can extract
a relatively low-amplitude input signal from a high-common-mode voltage
so the E Series DAQ device can measure the input signal. They also can
amplify and filter the input signal, resulting in higher measurement
resolution and accuracy.
SCC-AIXX modules are available in a range of gains from 0.2 to 200. They
contain lowpass filters of either 10 kHz or 4 Hz bandwidth. Table 1-1 gives
the gain and bandwidth for each module.
Table 1-1. SCC-AIXX Module Input/Output Range, Gain, and Bandwidth
Model Input Range Output Range Gain Bandwidth
SCC-AI01 ±42 V ±8.4 V 0.2 10 kHz
SCC-AI02 ±20 V ±10 V 0.5 10 kHz
SCC-AI03 ±10 V ±10 V 1 10 kHz
SCC-AI04 ±5 V ±10 V 2 10 kHz
SCC-AI05 ±1 V ±10 V 10 10 kHz
SCC-AI06 ±100 mV ±10 V 100 10 kHz
SCC-AI07 ±50 mV ±10 V 200 10 kHz
SCC-AI13 ±10 V ±10 V 1 4 Hz
SCC-AI14 ±5 V ±10 V 2 4 Hz
Vedere la pagina 0
1 2 3 4 5 6 ... 17 18

Sommario

Pagina 1 - Input Modules

USER GUIDESCC-AI Series Isolated Analog Input ModulesThe SCC-AI Series isolated analog input modules (SCC-AIXX) can extract a relatively low-amplitude

Pagina 2 - What You Need to Get Started

SCC-AI Series Isolated Analog Input Modules 10 ni.comTable 1-2 shows suggested values for some of the parameters. Use default values for parameters no

Pagina 3 - Unpacking the Module

© National Instruments Corporation 11 SCC-AI Series Isolated Analog Input ModulesScaling Voltage MeasurementsIf you configured the SCC-AIXX using MAX

Pagina 4 - Installing the Module

SCC-AI Series Isolated Analog Input Modules 12 ni.comFilter CharacteristicsNumber of poles ...3Rolloff...

Pagina 5 - Connecting the Input Signals

© National Instruments Corporation 13 SCC-AI Series Isolated Analog Input ModulesEnvironmentalOperating temperature... 0 to 5

Pagina 6

SCC-AI Series Isolated Analog Input Modules 14 ni.comCalibrating Gain and Offset ErrorsThe SCC-AIXX is calibrated at the factory before shipment. If y

Pagina 7 - Configuring the SCC System

© National Instruments Corporation 15 SCC-AI Series Isolated Analog Input Modules9. Adjust the trimpot labeled Gain located on the top of the SCC-AIXX

Pagina 8 - Traditional NI-DAQ

SCC-AI Series Isolated Analog Input Modules 16 ni.comSCC-AIXX Module Pin AssignmentsFigure 1-6 shows the I/O connector pins on the bottom of the modul

Pagina 9

© National Instruments Corporation 17 SCC-AI Series Isolated Analog Input Modules9 +6 V10 GND11 AGND12 REF5V13 +15 V14 –15 V15 —16 —17 —18 —19 —20 —Ta

Pagina 10

National Instruments™, NI™, ni.com™, and NI-DAQ™ are trademarks of National Instruments Corporation. Product and company names mentioned herein are tr

Pagina 11 - Specifications

SCC-AI Series Isolated Analog Input Modules 2 ni.comConventionsThe following conventions are used in this guide:<> Angle brackets that contain n

Pagina 12 - Maximum Working Voltage

© National Instruments Corporation 3 SCC-AI Series Isolated Analog Input Modules❑ One or more SCC-SGXX❑ SCC-SG Series Strain-Gauge Modules User Guide❑

Pagina 13 - CE Compliance

SCC-AI Series Isolated Analog Input Modules 4 ni.comInstalling the ModuleCaution Refer to the Read Me First: Safety and Radio-Frequency Interference d

Pagina 14

© National Instruments Corporation 5 SCC-AI Series Isolated Analog Input ModulesConnecting the Input SignalsThe SCC-AIXX has a fixed screw-terminal re

Pagina 15 - . Turn the trimpot

SCC-AI Series Isolated Analog Input Modules 6 ni.com Figure 1-3. SCC-AIXX Signal ConnectionsThe inputs are designed in a floating (nonreferenced) sin

Pagina 16

© National Instruments Corporation 7 SCC-AI Series Isolated Analog Input ModulesIf a high common-mode voltage is present, connect the negative input p

Pagina 17

SCC-AI Series Isolated Analog Input Modules 8 ni.com4. In the SCC Connector Block Configuration window, configure the SCC system as follows:a. Specify

Pagina 18 - *371066A-01*

© National Instruments Corporation 9 SCC-AI Series Isolated Analog Input Modules10. If you make a selection mistake, select the socket and click Remov

Commenti su questo manuale

Nessun commento